TRUNG TÂM DỊCH VỤ PHÂN TÍCH THÍ NGHIỆM THÀNH PHĨ HỊ CHÍ MINH

(CASE)

TAP HUAN

TAP HUAN KY THUAT PHAN TICH

QUANG PHO PHAT XA NGUYEN TU CAO TAN PLASMA ICP-OES

Trang 3

By đồ" vđu Trương

Inductively Coupled Plasma Theory đt CASE

| KHUNG i 2

What Applications is the ICP Used For? a8 CASE

¢ Geochemical Exploration

¢ Metallurgical Process Control

e Wear Metals in Oils

¢ Environmental (Waters, Soils etc)

¢ Foods

¢ Agricultural

Trang 4

"Bohr Model of Bee a8 PAGE

ee a)

> Ground State Electron (11 °vử9

> Quantized Orbitals

Electron returns ~~ Origins of Atomic Spectra 3j8 CASE

to lower energy level Decay with

Emission of Light - ata specific = wavelength Z” Excited States

>» Many transitions are possible AE = hv = he/A

> Emitted light depends on energy difference of

_two states

Trang 6

ñj$ CASE

Spectroscopic Information

> Qualitative

> related to wavelengths at which radiation is emitted or absorbed

> Quantitative

> related to amount of radiation emitted or

absorbed at analyte wavelength » depends on concentration

Quantitation in Atomic Emission & CASE

Concentration proportional to intensity

emitted at the analyte wavelength

Standard

Emission

Trang 7

i aj8 CASE

‘What is a Plasma?

>A plasma is a stream of highly ionized gas containing an equal number of electrons

and positive ions

> A plasma i is electrically conduetive re eee an

> A plasma is affected by a magnetic field 7⁄“”- “”⁄⁄

ñÿậ CASE

Inductively Coupled Plasma

(ICP)

Trang 8

ICP Characteristics - ae CASE :

> Low Background @ %&° G9 Ge AA! bate greued cam The

» Low Detection Limits for 70 Elements (Typically

low ppb)

» Inherently multielement

>» High sample throughput

» Wide Calibration Range

> Major, minor and trace in one solution

>» Few Chemical Interferences ,

a8 CASE

Plasma Background

>» The plasma emits a continuum of radiation

> The intensity increases two orders of magnitude between 200 and 400 nm >» Wavelengths below 190 nm require

Trang 9

Temperature (K) £ 10% 6000 6200 6500 6800 excitation {ion} mit M°" ionization {| excitation * _ (atom) M “hy atomization tl (ga) MX vaporization † (solid) desolvation (solution)

Figure 2-3 Zones of the ICP IR - Induction Re-

gion, PHZ - Preheating Zone, IRZ - Initial Radia- ion Zone, NAZ - Normal Analytical Zone

Trang 10

How is a Plasma Formed? aR CASE Ignition Sequence | 3{8 CASE

Argon Plasma Excitation Energy

In general, transitions observed in an ICP are

limited to those of less energy than the ionization energy of the gas

Ar —— Art +e 15.4 eV

Trang 11

1A

Optical

Trang 12

PHZ IRZ NAZ Tail Plume

ie r Gas Position PHZ (Pre Heating Zone) O Tail Plume

IRZ (Initial Radiation Zone) Decay to Ground State al Analytical Zone)

ntly tonic Emission SELF ABSORPTION of lic

zone for excited states

338 CASE

Tail plume must be removed for Axial viewing

Shear gas design

Extends working range

No maintenance necessary

No consumables to replace

_ Shear gas of air-not expensive AT :

a gemminates carryover anc

Trang 14

ICP Spectroscopy Basics 7 é{8 CASE Spectroscop) KH een Diên ——> Ph — > 02ppm si As —> 0.1ppm

ICP is an argon plasma maintained ~ OES is spectrometer which

by the interaction of an RF field separate and measure the

and ionized argon gas (~6000 - ; ,

: emitted light

10000k)

=

Convert elements in sample to atoms Wavelengths are characteristic of

* Excites atoms to emit the light element

Trang 15

Nebulizer Injector

BH Chamber met ue!

ee

Sample Peristaltic Pump

Major Components of Optima 5X00, 7X00, 8X00

` SPECTROMETER Sample SAMPLE INTRODUCTION ise" Anal RF Controller | _d 3000 Wats `

Power Supply COMPUTER, WINLAB32 S/W

RF GENERATOR as Sample Insoduetion ‘Compartment with

Trang 16

Cancels 3t HE) tì = =

5 KR usr?

SSS uULUUTT oyStem ae

Two Options - a ae CASE

\ Torch Coupler

Beary plan, Nebulizer Boos yon

“Spray Chamber

7 Clamp fe 3ẽ=——

Tubing Coupler ania

ORES)

Spray

Chamber

Tygon Tubing

ta tị

Scott (Double Pass) Spray Chamber with

aS iS

ni _ i

Figure $19 —inetaiting the cycionic epray chamber,

th Biome — Tá, Nếu ise Flow nh In, (Gem Tips) _

Trang 17

New Plasma Generation Technology

Patented Flat Plate™ Plasma Technology

Generates a unique symmetrical plasma

Same robust, matrix-tolerant plasma is generated

Almost half the argon consumption of helical load-coil systems

( 8L/min argon as compared to 15L/min ) Maintenance-free and requiring no cooling

Trang 18

ICP-OES OpticalSystem — _ ẩj$ CASE -

The optical system used for ICP-OES consists Of a spectrometer that is used to separate the

individual wavelengths of light and focus the desired wavelengths ậ E fr onto the detector

a CN J

Z z2 ot

» Sequential or Scanning ICP-OES(a fing, ng f8 - a ES ae a Se _ = wo “

n

* select any wavelength and focus it ona single detector

* design with ne- moving grating resulting

: f simultaneous measurement

® Measure one element at a time by moving ;

grating, which can lead to longer analysis

times 3 - Echelle Grating Parabola th | eis rea) ĩ | ee Disnaraat: Pe US ieee 0i SS | Pe > 1 i & oe Detector ị | L E——~——— @ộ@ ` Ấm: & BI = | fi z 5) | ý Flattener

Intermediate Silt Flier baie for

| 1⁄2 ma

——— the neon reference beam

iu 1 O) 01 TT TP ; Canna) =1) PS LÍ belies == : eed ICP-OES Detector : đi CASE cbcfenb Ay

» Charge-Coupled Device (CCD) Detector — detector use for Sequential ICP-OES » Segmented-array Charge-coupled

Device (SCD) Detector — the detector use :

for Simultaneous \cp- cee gio be a

Đ

<= Cross-Disperser Grating Dispersion ns

Register for the it

raterenioe trepauretint High Tis;

là :

2mm CCD array J

.” s

Wavelength

a

within the Order |&

° ý Ẹ s 3.5 mm CCD array ũ Low 7 ĩ ụ $ 130 120 110 100 80 80 70 $0

High Order Number Low Register for the 1870m

403nm

Output

analytical measurement _LoW `

Trang 20

meyer ne

Avio and Optima Series Product Line _ Wlz.4f

7 5c rete ee Berean

: D902 :E : b4 he

Optima Hybrid) FcheleDouble ˆcCD 160-900nm

8000 DV imultaneous Monochromator

Avio200 - Hybrid Echelle Double ccD 160-900 nm Dual View No _ Simultaneous Monochromator

Optima True ˆ”ˆ Echalle | Dual sco 163-782nm Radiai j Yes

7300V Simultaneous Polychromator / View - :

Optima True Echelle Dual SCD 163-782 nm Dual View Yes 8300 DV | Simultaneous Polychromator

Introduction

>» Models

* Optima 8000 —Avio 200 (Sequential), Optima 8300 — Avio 500 (Simultaneous)

« The Optima 7000 DV is replaced by

ñj$ CASE the Optima 8000

* The Optima 7300 DVis replaced by the Optima 8300 >» Changes

New Flat Plate™ RF Generator Desi

Syngistix software ign, new PlasmaCam"™ ,ICP WinLab32 software or Qqverg

Trang 22

Features Designed for Customer Na

Flat Plate Plasma

Technology

New Plasma Generation Technology

>

> >

338 CASE

Color Plasma Cam

| Syngistix ICP

Version 2.0

338 CASE

Optima 8300/7000 AVIO 00/200

Patented Hat “Induction Plate” Design

Generates a unique symmetrical plasma

Optimal performance at reduced argon flow while preserving stability and precision

Reduced cost of ownership (half the argon

consumption)

Long-life aluminum plates do not require cooling Ease of Manufacture and Service

Trang 23

oe ij§ CASE

Who Should Consider Avio 200?

The Avio 200 » Upgrade/replacement AA

Moderate to low sample load market ee

Typically less than 100 samples a day » Lower speed of analysis 1 to 5 elements per sample

Best analytical performance Small- to medium-size labs

Low to moderate sample loads Complements , Ỷ COD

snj ecto" phew Gy cokh GEMS 75 Gin WP dU Sent Ve) Wek Gud

338 CASE

Avio 500

The Avio 500

> Challenging analytical problems > Demands sample throughput > Ultimate analytical performance > Performance and productivity

» Simultaneous measurements: Speed » Better results than the competition

> Superior optical system > Superior resolution — > Superior software

“Tend to be large companies with moderate to heavy sample loads and want the

best product, service and support

Trang 25

than Th, Phubhg Ngoc |

3B CASE

mat Thay oh hes wtp phar 24

A z man

I/PHẦN TÍCH NUTRITION (Na, Ca, Fe) TRONG THUC PHAM BANG

PHUONG PHAP QUANG PHO PHAT XA PLASMA (ICP- OES)

oe j đŸ§ CASE NỘI DUNG CHÍNH

=1 Giới Điệu thành phần dinh dưỡng s2, ae trinh phan tich mau :

3 Phan tich Na, Ca, Fe trén lcP -OES_

4 Cơng thức tính giá trị dinh tường ( Daily

value)

Trang 26

eo: f afi CASE

I KHAI NIEM VE NUTRITION

- Thực phẩm đóng một vai tro căn bản trong việc cung cấp nguồn năng lượng sống cho cơ thể

- Thực phẩm được cấu tạo bởi các chất bổ dưỡng (Nutrients), bao gồm những thành phần hóa học,để ni sống cơ thể

- Dinh Dưỡng : hấp thu những chất bổ dưỡng

trong thức ăn để nuôi dưỡng thân thể

(goi“Nutrition”)

ee af CASE

I KHAI NIEM VE NUTRITION

- Thyc pham được phân thành các nhóm tùy vào thành phan dinh dưỡng nhiêu nhất

Trang 28

Vai trò của Natri (Na- Sodium) iy CASE

- Khống chất có vai trị thiết yếu đối với cơ thể con

người

- Chất điện giải quan trọng bên cạnh Kali, Canxi và Magie

- Natri vào cơ thé chủ yếu là trong muối ăn - Natri clorua

(NaCl)

- Nếu lượng Natri trong cơ thể quá cao (với những người ăn mặn) sẽ gây ra nhiều tác hại cho sức khỏe như tim đập nhanh, mắt cân bằng điện giải, mỏi cơ, co giật, gây hại cho gan và thận, thậm chí tử vong

Vai trò của Canxi (Ca - Calcium)_ 8 CASE

Giúp cơ thê hình thành hệ xương và răng vững chắc

Đảm bảo chức năng thần kinh và sự đơng máu bình thường

Ca trong máu giảm -> cơ thể lấy Ca từ xương dé tham gia

các quá trình chuyên hóa >3 đau nhức các xương đặc biệt các xương dài ở trẻ đang tuổi phát triển; -> tình trạng mat

ngủ, tính tình nóng nảy

Thiếu Ca lâu dài trong khẩu phần > bệnh cao huyết áp và

ung thư ruột

Lượng Ca ăn vào dư thừa -> Ca được bài tiết ra khỏi cơ thể Dùng thuốc Ga liều cao, kéo dài > soi than, canxi máu cao,

thận làm việc kém hiệu quả ; giam hap thu cac chat khoang

cần thiết khác (ví dụ: sat, kém, magié, phosphor, iod, dong)

Trang 30

Nhu cầu Sắt của con người ẩŸš CASE

Nhóm giới tính, Nhóm giới tính,

độ tuổi( năm ) RDA( mE) | gã tuổi năm } EDEtee)

Em bé ( 05-1) 11mg Phụ nữ có thai 27mg Trẻ nhỏ IPhụ nữ 1-3 7mg 19-30 18mg 4-6 10mg 31-50 1amg 7-9 10mg 51-65 10mg 66-75 1omg >75 10mg

Con gái Đàn ông

10-12 10-15mg 19-30 10mg

13-15 15mg 31-50 10mg 16-18 15mg 51-65 '10rng

Con trai 66-75 tome

10-12 10mg >75 10mg 13-15 10mg 16-18 12mg mỗi ngày

Nhu cầu Natri của con người

Tổ chức Y tế Thế giới (WHO) khuyến nghị ít hơn 2 gram natri - Hiệp hội Tim mạch Mỹ đưa ra khuyến nghị thận trọng hơn là 1,5 gram natri mỗi ngày, đặc biệt đối với những người có nguy cơ mắc bệnh tim :

ayo CASE

Trang 32

ao CASE SƠ ĐÒ XỬ LÝ MẪU

Đồng Than hóa Để nguội -

nhất Nung 525 + Tro hóa, nung

mau, can : 25°C, 3-5h Xử lý lại mẫu —”{ lại > tro trang :

3-5(9) bang HNO3 (1:)1 |

Do ICP - Định mức _| Hoda tan tro = OES 50 mL 10mL HNO3 1M "mm iŸ§ CASE

III PHÂN TÍCH MẪU TRÊN ICP-OES

Bước 1 :Lập chương trình - Tạo method

- Chọn nguyên tố cần phân tích, chọn bước sóng thích hợp

(Ca A=317.933nm , Na A=589,592 nm, Fe A=238,204 nm)

Trang 33

é i 3]8 CASE

Thông số kỹ thuật của máy ICP-OES

PerkinElmer 5300DV

Plasma Set Plasma (L/min) :16

Aux (Limin) :0.3

Neb (L/min) :0.9

Power (W) 71350

_ View Dist 215 ae Ot

Ca, Na : Radial (xuyên tâm)->!⁄F9 = + Fe : Axial (doc truc)> "aq Te me -

i٤ CASE

Thông số kỹ thuật của máy ICP-OES

PerkinElmer 5300DV

"ST pump: HỆ Sample Flow Rate (mL/min) :1.5

Flush time (sec) : 15

Trang 34

Ai os 38 CASE

PHAN TICH MAU TREN ICP-OES

Bước 2: Xây dựng đường chuẩn (Define standards)

Lập đường chuẩn phù hợp với mẫu đo:

‹- Ca ; 5-100 ppm (Khoảng tuyến tính: 1-200ppm, R=0.9997)

s Na : 5-100 ppm (Khoảng tuyến tính: 1-200ppm, R?=0.9997) » Fe :0.1 - 5ppm (Khoảng tuyến tính :0.1-25ppm, R2=0.9998)

PHAN TÍCH MAU TREN ICP-OES aie CASE

'Bước 3 : phân tích mẫu

a Do blank

b Do chuan c Do mau

Chú ý:

- Mẫu được chuyển về dạng dung dịch khơng có cặn

lơ lửng

Trang 36

ee ; : đÏ$ CASE

_IV.TÍNH TỐN KÉT Q

s Tính tốn hàm lượng các nguyên tố cần phân tích : trong mẫu thực tế ban đầu

: : HL: hàm lượng ngun tơ

có trong mẫu phân tích -_G: lượng can V: thể tích định mức f: hệ số pha loãng re : 33 CASE

TAI LIEU THAM KHAO :

1 AOAC 985.35

2.Trích dẫn từ nguồn của Healthvn

3 Trích dẫn từ nguồn của WHO

4 Trích dẫn từ nguồn Viện Dinh Dưỡng Quốc Gia

Trang 38

348 CASE Các phương pháp phân tích Pb, Cd

* Quang phé hap thu nguyen tử Quang phổ phát xa nguyên tử Phổ khối lượng ngun tử

Phương pháp vơn-ampe hồ tan

3j8 CASE

Hàm lượng cho phép của Pb, Cd trong một số nền mẫu ˆ Giới hạn (m8/ke}

Thực phẩm - ni

Heo, Bò, Gà cee 01 U05

Cá ae ; ‘ 0.3 0.4

Quả 0.1 0.05

lãi Tố sử 08 - 0.2

Trang 40

pene gar aE eG PSM Ee

3f8 CASE

xs

- CÁC BƯỚC CHUAN B] MAU

° Đồng nhất mẫu

* V6 co’ hoa mau

° Hoà tan và định mức a8 CASE ĐỒNG NHẤT MẪU

Mẫu được cắt nhỏ, trộn đều, xay nhuyễn bằng các dụng cụ thích hợp