Trang 1EXPERIMENTAL REPORT
Instructer: Dr Nguyen Vinh Tien
Subject: Practice of food analysis
ID Class: PRFA414550E_22_2_05FIE
Group: 03
Trang 2Assessment of Accuracy and Precision
1 Introduction
Determine the deviation/error of mass (analytical balance) or volumetric instrument(burette, pipette, micropipette, volumetric flask, beaker) From there determine the errorrange of the instrument in the experiment, estimate the error of the experimental results.Thereby making comments and comparing the deviations between supplies
2 Objective
Assess accuracy and precision of data generated in using balance, mechanical pipettes and volumetric glassware
3 Principle
Using equipment in analytical tests helps ensure more accurate and precise results
4 Supplies and equipments
Trang 35 Procedure
5.1: Evaluate the error of the pipette
1.Put the 100mL beaker on the analytical balance
2.Press Tare to not calculate the beaker weight, reduce to the value 0.00g
3.Use a 100mL pipette to accurately draw 100mL of distilled water into the beaker placed on the balance
4.Weigh the volume of water added
Note: It is not necessary to pour the water from the previous measurement out of the cup, just press Tare and then take another 1mL of water
(Do the same with the 1mL Pipette)
Repeat the experiment 6 times
5.2: Evaluate the error of the buret
1.Place the 100mL or 20mL beaker on a precision balance
2.Press Tare to subtract the cup weight, reducing it to a value of 0.00g
3.Using a burette (25/50ml) add 10ml of distilled water (ml) to the beaker
4.Drain 10 mL from the burette in turn until the amount of water in the burette is not enough 10
mL, then add water and adjust to zero
5.Repeat until 6 measurements are obtained
Then repeat the process as above but take 1 mL from the burette
5.3: Evaluate the error of the micropipette
1.Place the 100mL or 20mL beaker on a precision balance
2.Press Tare to subtract the cup weight, reducing it to a value of 0.00g
3.Using a 1000 µL micropipette add 1ml of distilled water (ml) to the beaker
4.Weigh the volume of water added
Repeat this process 6 times
Note: It is not necessary to pour the water from the previous measurement out of the cup, just
Trang 4press Tare and then take another 1mL of water.
5.4:Total amount of water and amount of water taken from the volumetric flask1.Place the 100ml/Beaker 100ml volumetric flask on the precision balance
2.Press tare to subtract the cup weight, convert to the value 0.00g
3.Fill the flask up to the mark with water and weigh the amount of water in it
4.Then tare a 100 mL beaker and pour water from the upper volumetric flask into the beaker, weighing the amount of water removed from the flask
Repeat this process 6 times
5.5: Accuracy of measurements from balance
1.Press the balance tare to return to the 0.00g value
2.Put weights 50g, 100g, 200g respectively on a precision balance.Record the result
Note:Do not touch the weight with bare hands to avoid the oils from sweat sticking the weight.Work with precision balances and analytical balances Then record the number of the balances
6 Calculation and results
6.1:Evaluation of pipette error
Volumetricpipette1
ml
10ml
e
e
Trang 59,9292 0,04±24n=6
9,9230 0,01±56
6.2:Evaluation of Buret 25ml error
Buret1
ml
10ml
Trang 76.3:Evaluation of micropipette error
Mechanicalpipettor
Trang 86.4:Total amount of water and amount of water taken from the volumetric flask
Mass in analytical balances(g)
Errors may appear due to: Instruments are relative, misaligned, chipped, sweaty hands
Trang 9When taking water into the cup and weighing it, there may be some water left on the wall of the machine
The person conducting the experiment did not perform the operation correctly
Due to laboratory conditions such as when the balance meets the wind, it will affect the weighing results
Because there are computational differences between the data, it is necessary to perform many times to obtain a set of values around the true value, minimizing errors
In addition, it is necessary to limit some human factors and surrounding factors when
conducting experiments
8 Questions
1.In theory, how the standard deviation, coefficient vvariation, mean, relative error, and 95% confidence intervals are affected by: (1) multiple replicates and (2) measurements bigger? When measured with pipettes and burettes, is this reflected in the results of
experiments with n = 3 vs n = 6, with 1 mL vs 10 mL?
In general, the standard deviation, coefficient variation, mean, relative error, and 95% confidence interval will be less than or equal to when the experiment is repeated many times, thereby reducing the difference between the samples calculated value This is also shown by thedata obtained during the experiment when calculating the values with n = 3 and n = 6
2.Why should relative error and coefficient variation be used to compare the accuracy andprecision of the 1 mL and 10 mL volumes obtained from the pipette and burette in sections
2 and 3, instead of just taking the average value? and standard deviation?
Instead of just the mean and standard deviation, we use the relative error and the coefficient of variation to compare the accuracy and precision of the 1mL and 10mL volumes taken from the pipette and the burette for the coefficient of variation, which is more reliable, in addition to showing more clearly the range of errors between the values, the coefficient of variation also helps to gather data in a region and the acceptability of the obtained value
4.If the accuracy and/or precision of the micropipette is lower than expected, what can be done to improve these characteristics?
Pre-wash the tip: Pre-wash the tip to ensure that all internal components are in contact with the
Trang 10solution Aspirate the sample into the tip and then back-flush or flush the waste cup to rinse the tip, increasing the accuracy of the pipetting process.
Reverse pipetting technique: This technique is used when aspirating viscous solutions (blood samples, ) or volatile solvents This technique also allows aspiration of microscopic volumes such as 0.5 µl or less The reverse pipetting technique begins with pressing the plunger down to the stop position 2nd, thus including an additional volume remaining during sample aspiration, not included in the final discharge
Consider ambient temperature: People or companies often align pipettes at room temperature,
so if you use a pipette at a different temperature (e.g in a cold room), you may not get the correct volume corpse
Consider sample temperature: When continuously aspirating samples at cold temperatures, the volume of solution at the first flush is always greater than the volume aspirated, but not on subsequent aspiration with the same tip for exact volume Therefore, it is recommended to rinsethe first aspiration into the solution tube and start aspirating from the second time
Use volume-matched micropipettes: Micropipette accuracy will decrease when the discharge volume is close to the smallest volume the pipette can handle
Install the pipette:
Adjust more than 1/3 of the turn to the desired volume Then adjust back to the required volume
Do the same with different volumes, with different times
5 In a titration experiment with a burette, should the stopping volume be too much less than 10 mL? Is it better to use 10 mL burette or 50 mL burette for accuracy and precision?Why?
The stop-point volume should not be too much less than 10mL because the opening of the off valve is unstable, which can lead to errors in titration or reading Using a 10 mL burette offers greater accuracy and precision Because the smaller the error of the device, the higher the accuracy and precision Where the 10 mL burette has an error of ±0.02 (mL), while the 50 mL burette has an error of 0.05(mL)
shut-6 Is the volume indicated on the volumetric flask a “take in” or “take out” volume? What does the volume on the pipette indicate?
The volume of each volumetric flask is the volume contained to exactly a given volume The volume indicated on the pipette is the volume aspirated to an exact volume within a given
Trang 118 What sources of error (human or tool; systematic or random) have or might have
appeared in sections 2-4, and how can they be eliminated or reduced?
In sections 2 and 4 there are error sources:
System error
Random number
Errors due to people and tools
To eliminate and minimize errors:
Repeat the measurement several times
Test and calibrate measuring devices and instruments
Trang 12Determination of Moisture & Ash Content
Trang 39Determination of reducing sugar by colorimetric spectrophotometric method with DNS reagents
1 Introduction
The spectroscopic method is based on the changes of electromagnetic radiation as
it interacts with matter When monochromatic light passes through a transparentcolored solution, its intensity is reduced, depending on its nature and soluteconcentration in solution In a given concentration range, this decrease (moreprecisely the log of the decrease in intensity, or optical density of the solution) isproportional to the concentration of the solute causing the color That is the basis
of the Beer - Lambert law that helps determine the concentration of solutes in asolution
Trang 40the compound formed is proportional to the concentration of reducing sugars within a certain range
Based on the standard curve equation will determine the reducing sugar content of the test sample
4 Chemicals, supplies and equipments
Chemicals & food sample
● DNS reagents included:
★ Distilled water: 60 – 70mL
★ Solid NaOH: 1.6g
★ DNS (C7H4N2O7): 1g
★ (KNaC4H O 4 4.4H2O) salt: 30g
➔ Stir the mixture completely with a magnetic stirrer and make up to the 100mL mark with a volumetric flask Reagents are stored at low temperature of 2 – 8oC and without light (solution is good for 10-15 days after reconstitution)
● Standard Glucose solution 0.1%
● Distilled water
● Coca & pepsi
Equipments
Test tubes
Trang 41500 mL)Standard Glucose
Trang 42Block diagram 5.1 How to Determination of reducing sugar by colorimetric
spectrophotometric method with DNS reagents
6 Results and Calculation
100 mL)Standard Glucose
-Table 5.2 The results were obtained after measuring the optical density at 540
nm using a colorimetric spectrophotometer6.2 Calculation
Trang 43When measured on a 540 nm spectrophotometer:
➔ There is 6.875 g/100 mL of reducing sugar in the coke sample
Based on the 100ml nutritional value table in a coke bottle The amount of sugar in 100
mL of coke is 10.6g
The amount of reducing sugar in the total carbohydrates in coke is:
% Reducing sugar = x 100 = 64.858%
Trang 44➔ There is 5.284 g/100 mL of reducing sugar in the Pepsi sample
Based on the 100ml nutritional value table in a Pepsi bottle The amount of sugar in 100
- The correlation coefficient R = 0.9827 is close to 1, so the experimental values havelittle difference compared to the standard curve
The cause of the error
- Incorrect sample dilution
- Prepare chemicals for non-standard experiments
- Incorrect chemical collection
- Cuvet is scratched, dirty, so it affects the measurement result
- The operation is not standard
Methods of minimizing errors
- Accurate sampling
- Check and clean tools before taking chemicals and doing experiments
- Performed under more secure experimental conditions
- Make sure cuvettes are clean and not scratched
Trang 45- Repeat the experiment many times to come to the most accurate conclusion
- When turned on, it is recommended to wait 10-15 minutes for the spectrometer to stabilize before testing
Determination of solution protein by Biuret method
1 Introduction
The sample of solid or liquid food contains a certain amount of soluble and insoluble protein Besides that, samples of food also contain a certain amount of non-protein impurities that contain nitrogen, which is calculated by using Kjeldahl or Dumas method
2 Objective
However, the compound contains non-protein nitrogen due to no peptide circuit so it doesnot create complexes with Cu2+ ions Therefore, calculating protein concentration will becorrect when determined by creating complexes with Cu2+ indicator And Biuret method
is one of the methods that help identifying soluble protein in food based on its ability in creating the complexes of peptide circuits with Cu2+ ions
3 Principle
The Biuret method is based on the complexification of Cu2+ to functional groups in the protein’s peptide bonds The formation of a Cu2+ protein complex requires two peptide bonds and produces a violet-colored chelate product which is measured by absorption spectroscopy at 540 nm Over a given concentration range, the measured absorption at
540 nm is linear with respect to the concentration of total protein This relationship will allow us to create a standard curve that is used to calculate the concentration of an unknown sample
Trang 464 Chemicals, supplies, and equipments
a) Supplies and equipment:
5.1 Prepare biuret reagent
Figure 5.1 Diagram prepare biuret reagent5.2 Prepare standard protein solutions
Trang 475.3 Build standard curve
Proteinconcentration
(mg/ml)
Emission at540nm
Trang 4810 0.292
5.3 Determine the absorbance of food samples
6 Data and calculation
The emission of Peptone solution is 0.113 nm
Following the equation from standard curve, we have concentration of protein in peptone solution:
Trang 497 Discussion
Through this experiment, we can see the concentration of soluble protein in peptone solution is 3.396mg/ml
a/ The cause of the error
- The sample weighing may not be entirely accurate due to external environmental factors or insufficiently clean sample preparation equipment
- The cuvette may not have been thoroughly cleaned, or there may be residues during handling and transportation
- The storage of the Biuret reagent is not strictly controlled
- Minimising the contact of the Biuret reagent with air and light
Determination of total lipid in solid samples by
Soxhlet method
1 Introduction
More than 95% of lipids in food are esters of high molecular fatty acids and glycerol Thecommon lipids are vegetable oil (from soy, peanuts, olive, …) and animal fat (milk,
Trang 50brain, … specially adipose tissue)
2 Objective
Determination of lipids in solid food samples by extraction with organic
solvents( Soxhlet method)
3 Principle
Fat is extracted with organic solvent which is heated and volatilised then condensed above the sample Solvent drips onto the sample, soaks it to extract the fat Final fat content is measured by weight loss of the sample
4 Chemicals, supplies, and equipments
a) Supplies and equipment
Trang 51Place the sample (wrapped in paper) into the extraction cylinder The solvent is then poured from the top of the condenser through the extraction cylinder and finally down to the solvent flask Turn on the water to cool the condenser and turn on the heat source to evaporate the solvent The volatile solvent will be cooled by a condenser and condenser tube system, flowing down to the extraction cylinder and the solvent tank.
6 Data and calculations
7.1 The factors affect the result
Weighing samples is not very accurate because it can be influenced by experimental factors or because the sample preparation equipment is not very clean throughout the weighing process
The sample's solvent or moisture cannot be entirely eliminated by drying