Put scanner from middle line of axillary toward under and from outside toward inside breast, flatten breast linear and slightly push under and inside. Rotate scanner in order to[r]

(1)(2)

2 CONTENT

1. Breast anatomy

2 I di i d i di i f 3D b l d

2. Indication and contraindication of 3D breast ultrasound 3. Perform technique

4. Image assessment

Breast anatomy

(3)

Indication and contraindication

1 Indication

• Combine with mammogram for breast cancer screening every year • Evaluating development process of linear

• Linear with high intensity on mammogram image

• Women with abnormal symptom in linear Phụ nữ có triệu chứng bất thường tuyến vú

2 Contraindication

• Don’t have absolute contraindication Althought, shouldn’t perform in case: - Case don’t have capability of withstand compress like paint and breast abcess Các trường hợp khơng có khả chịu đựng đè ép cảm giác đau có áp xe vú - Sau phẫu thuật thẩm mỹ tuyến vú với chất độn lớn cứng (rất khó ép khó

đánh giá tổn thương) - Vú nhỏ người gầy - Có vết thương hở vú

Prepair equipment

• Invenia Automatic Breast Ultrasound System • Ultrasound gel : 250mlg

(4)

4 Ultrasound system

Install stabilized film

(5)

Prepair patient

This technique don’t demand special prepairation

• Double check patients (Name, DOB,….) Ask information related to previous breast exam Ask about cause, position of pain and how they feel

h k

at the moment, reason to take exam • Let patient change to hospital cape • Explain about procedure :

+Number of position on each side(3 position per side minimum) +Take about 20mins to perform

+ Pressure feeling on breast Patients should relax and talking, moving should be restricted during the scanner moving

be restricted during the scanner moving

+ Access size of breast by eye: Small(S)- Medium (M)- Lagre(L)

Patients have peacemaker and other medical implant should ask doctor opinion before start

System screen

• Fill up patients information: Name, information: Name, DOB,PID

h b i

(6)

6 Patients position

• Patients lay face up on scanning table, put a cushion under patients head in order to spread mammary gland spread evenly

• Nipple point up and at the central of mammary gland • Raise patients have above their head

• Patients head turn to not scanning side

(7)

Gel applying technique

(8)

8 AP view

• Decribe scanning area

‐ Have middle mammary gland: included nipple at central.Have middle mammary gland: included nipple at central

Scanner postion

• Nipple pointing arrow in middle of scanner in corresponding position with

AP view

corresponding position with nipple

• Scanner position at lower edge of nipple

• Put pressing table from above to put pressure on mammary gland in order to flatten mammary gland

(9)

Lateral view

• Describe scanning area:

Lateral linear: include linear from Lateral linear: include linear from middle axillary line and lateral edge of breast Nipple will be in inner lower conner of stabilized film

Lateral view

Scanner position: Scanner position:

Scanner move toward axillary and outside breast.

Put scanner from middle line of axillary toward under and from outside toward inside breast, flatten breast linear and slightly push under and inside Rotate scanner in order to

k b d id

take above and outside must guarantee contact with outer edge of breast

(10)

10

Medial view

Mô tả vùng thăm khám:

• Nhu mơ vú trong: Bao gồm nhu mơ vú từ đường xương nhu mô vú từ đường xương ức qua nếp nằn vú 2cm Núm vú phía biểu tượng mũi tên định vị

Medial view

Scanner position

• Technician move to opposite side and scanner move along g

• Place inner edge of scanner straight in middle line of sternum and take 2cm above breast fold Put pressure and push scanner toward outer edge of breast but guarantee scanner contact with mammary gland ( don’t have space for air to come in)

Ni l t f d • Nipple at upper area of scanner and

right above nipple positioning arrow • ush press button depend on how much

patients can handle

(11)

Additional position (INF, SUP, UOQ)

Meaning of additional position

• INF position: additional for missing part under breast fold.

• SUP position : additional for missing part in above( if breast is too big and mammary fill up above) • UOQ position: additional to axillary tail (if patients

(12)

12 ultrasound

(13)

Image asscessment

• All mammary gland need to be expose according to scan position Guarantee to take all mammary gland in all direction. position Guarantee to take all mammary gland in all direction • Nipple at middle position in AP position, at lower under

conner in LAT position, at upper middle in MED position (compare to nipple positioning arrow)

• Image don’t be affected by motion artifact or air bubble

(14)

14 • REFERENCE DOCUMENTS

Invenia ABUS scan Station Positioning Guide  (GE Healthcare).