Comparison of the SSSperm testing kit with the Halosperm testing kit in an analysis of sperm DNA fragmentation

6 29 0

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Định dạng
Số trang 6
Dung lượng 0,98 MB

Nội dung

The objectives of this work were to accurately evaluate testing kits that analyse sperm DNA fragmentation in infertile men and to provide a comparison of an improved testing kit (SSSperm testing kit) to the existing Halosperm testing kit in an analysis of sperm DNA fragmentation.

Life Sciences | Medicine Doi: 10.31276/VJSTE.62(2).71-76 Comparison of the SSSperm testing kit with the Halosperm testing kit in an analysis of sperm DNA fragmentation Le Thi Quyen, Nguyen Thi Minh Thu, Le Thi Minh Phuong, Nguyen Thi Trang* Hanoi Medical University Received 18 April 2020; accepted 21 May 2020 Abstract: Introduction The objectives of this work were to accurately evaluate testing kits that analyse sperm DNA fragmentation in infertile men and to provide a comparison of an improved testing kit (SSSperm testing kit) to the existing Halosperm testing kit in an analysis of sperm DNA fragmentation A cross-sectional study was conducted on 300 semen samples from infertile men with a sperm concentration ≥1 million/ml using the Bland-Altman, T-test, and Pearson test for statistical study The SSSperm testing kit had a coefficient of variation of CV%=2.26%30% and had p=25% For ε, we selected Method of making templates: the test (using the SSSperm testing kit) was an sample size of 300 was2×0.25×(1-0.25)/(0.2×0.25) used 0.2 Therefore, n=1.96 avoided, which improvement of Fernandez, et al.'s SCD =288.12, procedurewashing (2005)was[10] using thecan lighten the halo colour Research design: a cross-sectional study Halosperm Halotech as follows: Methodkit offrom making templates: the test (using the SSSperm testing kit) was an Step Preparation of agar: an agarose Eppendorf tube was improvement of Fernandez, et al.'s SCD procedure (2005) [10] placed using into the the float and melt using a ofwater bath at 95-100°C for or in microwave for min, Vietnam Journal Science,as Halosperm kit from Halotech follows: june 2020 • Volume 62 Number 72 Technology and Engineering until Step it was completely melted semen samplestube were a PBS Preparation of agar: anThe agarose Eppendorf wasdiluted placed with into the solution that the concentration of sperm

Ngày đăng: 17/08/2020, 20:16

TỪ KHÓA LIÊN QUAN

TÀI LIỆU CÙNG NGƯỜI DÙNG

TÀI LIỆU LIÊN QUAN