Liên hệ 037.667.9506 hoặc email thekingheavengmail.com để nhờ đặt mua tất cả các tiêu chuẩn kỹ thuật quốc tế với giá rẻ. Tài liệu sẽ được gửi cho bạn trong 24 giờ kể từ ngày nhận thanh toán. Tiêu chuẩn ASTM là bộ tiêu chuẩn kĩ thuật cho các vật liệu và sản phẩm trên Thế giới được ban hành bởi tổ chức Quốc tế ASTM (American Society for Testing and Materials) – Hiệp Hội Vật Liệu Và Thử Nghiệm Hoa Kỳ. Những tiêu chuẩn mà tổ chức này đưa ra là sự đồng thuận và được cộng đồng quốc tế chấp nhận.
Trang 1Designation: D287−22
Standard Test Method for
API Gravity of Crude Petroleum and Petroleum Products
This standard is issued under the fixed designation D287; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision A number in parentheses indicates the year of last reapproval A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S Department of Defense.
1 Scope*
1.1 This test method covers the determination by means of
a glass hydrometer in conjunction with a series of calculations
of the API gravity of crude petroleum and petroleum products
normally handled as liquids and having a Reid vapor pressure
(Test Method D323) of 14.696 psi (101.325 kPa) or less
Values are determined at existing temperatures and corrected to
values at 60 °F (15.56 °C), or converted to values at 60 °F, by
means of Adjunct toD1250Standard Guide for the Use of the
Joint API and ASTM Adjunct for Temperature and Pressure
Volume Correction Factors for Generalized Crude Oils,
Re-fined Products, and Lubricating Oils (API MPMS Chapter
11.1) These tables are not applicable to nonhydrocarbons or
essentially pure hydrocarbons such as the aromatics
1.2 The initial values obtained are uncorrected hydrometer
readings and not density measurements Values are measured
on a hydrometer at either the reference temperature or at
another convenient temperature, and readings are corrected for
the meniscus effect, the thermal glass expansion effect,
alter-nate calibration temperature effects and to the reference
tem-perature by means of the petroleum measurement tables; values
obtained at other than the reference temperature being
hydrom-eter readings and not density measurements
1.3 The initial hydrometer readings determined shall be
recorded before performing any calculations Then the
calcu-lations required in Section 9 shall be performed and
docu-mented before using the final result in a subsequent calculation
procedure (measurement ticket calculation, meter factor
calculation, or base prover volume determination)
1.4 The values stated in inch-pound units are to be regarded
as standard The values given in parentheses are mathematical
conversions to SI units that are provided for information only and are not considered standard
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use It is the responsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.
For specific warning statement, see8.5
1.6 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for the Development of International Standards, Guides and Recom-mendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2 Referenced Documents
2.1 ASTM Standards:2
D323Test Method for Vapor Pressure of Petroleum Products (Reid Method)
D1250Guide for the Use of the Joint API and ASTM Adjunct for Temperature and Pressure Volume Correction Factors for Generalized Crude Oils, Refined Products, and Lubricating Oils: API MPMS Chapter 11.1
D1298Test Method for Density, Relative Density, or API Gravity of Crude Petroleum and Liquid Petroleum Prod-ucts by Hydrometer Method
D6822Test Method for Density, Relative Density, and API Gravity of Crude Petroleum and Liquid Petroleum Prod-ucts by Thermohydrometer Method
D7962Practice for Determination of Minimum Immersion Depth and Assessment of Temperature Sensor Measure-ment Drift
D8164Guide for Digital Contact Thermometers for Petro-leum Products, Liquid Fuels, and Lubricant Testing E1Specification for ASTM Liquid-in-Glass Thermometers E77Test Method for Inspection and Verification of Ther-mometers
1 This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products, Liquid Fuels, and Lubricants and the API Committee on
Petroleum Measurement, and is the direct responsibility of Subcommittee D02.02
/COMQ, the joint ASTM-API Committee on Hydrocarbon Measurement for
Custody Transfer (Joint ASTM-API).
Current edition approved Dec 1, 2022 Published February 2023 Originally
approved in 1928 Last previous edition approved in 2019 as D287–12b (2019).
DOI: 10.1520/D0287-22.
2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org For Annual Book of ASTM Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959 United States
Trang 2E100Specification for ASTM Hydrometers
E126Test Method for Inspection, Calibration, and
Verifica-tion of ASTM Hydrometers
E344Terminology Relating to Thermometry and
Hydrom-etry
E2251Specification for Liquid-in-Glass ASTM
Thermom-eters with Low-Hazard Precision Liquids
E2877Guide for Digital Contact Thermometers
2.2 API Standards:3
MPMS Chapter 9.1Test Method for Density, Relative
Density, or API Gravity of Crude Petroleum and Liquid
Petroleum Products by Hydrometer Method (ASTM Test
MethodD1298)
MPMSChapter 9.3Test Method for Density, Relative
Density, and API Gravity of Crude Petroleum and Liquid
Petroleum Products by Thermohydrometer Method
(ASTM Test MethodD6822)
MPMS Chapter 11.1Temperature and Pressure Volume
Cor-rection Factors for Generalized Crude Oils, Refined
Products, and Lubricating Oils (Adjunct to ASTMD1250)
2.3 ASTM Adjuncts:
Adjunct toD1250Standard Guide for the Use of the Joint
API and ASTM Adjunct for Temperature and Pressure
Volume Correction Factors for Generalized Crude Oils,
Refined Products, and Lubricating Oils: API MPMS
Chap-ter 11.1)4
3 Terminology
3.1 Definitions:
3.1.1 API gravity, n—a special function of relative density
60/60 °F (15.56/15.56 °C), represented by:
°API 5@141.5/~relative density 60/60 °F!#2 131.5 (1)
No statement of reference temperature is required, since
60 °F is included in the definition
3.1.2 hydrometer reading, n—the point on the hydrometer
scale at which the surface of the liquid cuts the scale
3.1.2.1 Discussion—In practice for transparent fluids this
can be readily determined by aligning the surface of the liquid
on both sides of the hydrometer and reading the Hydrometer
scale where these surface readings cut the scale (Hydrometer
Reading – Observed) For nontransparent fluids the point at
which the liquid surface cuts the Hydrometer scale cannot be
determined directly and requires a correction (Meniscus
Cor-rection) The value represented by the point (Meniscus
Read-ing) at which the liquid sample rises above the main surface of
the liquid subtracted from the value represented by where the
main surface of the liquid cuts the Hydrometer scale is the
amount of the correction or Meniscus correction This
menis-cus correction is documented and then subtracted from the
value represented by the Meniscus Reading to yield the
Hydrometer Reading corrected for the Meniscus (Hydrometer
Reading – Observed, Meniscus Corrected)
3.1.3 observed values, n—values observed at temperatures
other than the specified reference temperature; these values are only hydrometer readings and not density, relative density, or API gravity at those other temperatures
3.1.4 specific gravity, n—historical term, no longer used, which has been replaced by relative density.
3.2 Acronyms:
3.2.1 PRT—platinum resistance temperature device 3.2.1.1 Discussion—While there may be other types of
RTDs available, for Custody Transfer operations in conjunc-tion with other standards organizaconjunc-tions Platinum RTDs have been standardized on for their accuracy and discrimination
4 Summary of Test Method
4.1 This test method is based on the principle that the gravity of a liquid varies directly with the depth of immersion
of a body floating in it The floating body, which is graduated
by API gravity units in this test method, is called an API hydrometer
4.2 The API gravity is read by observing the freely floating ASTM hydrometer (calibrated for API gravity) and noting the graduation nearest to the apparent intersection of the horizontal plane surface of the liquid with the vertical scale of the hydrometer, after temperature equilibrium has been reached The temperature of the sample shall be read from a separate accurate ASTM thermometer placed in the sample, which meets either Specifications E1 or E2251 requirements or ASTM Digital Contact Thermometers, which meet Guide
E2877requirements The temperature determination device, be
it the bulb of a ASTM Thermometer (Specifications E1 or
E2251) or a sensor of a Digital Contact Thermometer (Guide
E2877) shall be placed at the same elevation (within the stated tolerances) as the hydrometer bulb
N OTE 1—Through various testings that Subcommittee D02.02 (Joint ASTM/API Subcommittee) measurement committee and others have conducted, it has been determined that temperature stratifications do exist vertically from top to bottom of a hydrocarbon container as well as across the diameter of the container Therefore, as temperature affects the viscosity as well as the fluid density, the buoyancy of the hydrometer floating in the liquid is therefore affected, clarifying procedures have been added to the Procedure section.
4.3 The observed hydrometer reading is corrected for the meniscus effect, the thermal glass expansion effect on the hydrometer, alternate calibration temperature effects and re-duced to the reference temperature by means of the petroleum measurement tables If necessary, the hydrometer cylinder and its contents are placed in a constant temperature bath to avoid excessive temperature variation during the test
5 Significance and Use
5.1 Accurate determination of the gravity of petroleum and its products is necessary for the conversion of measured volumes to volumes at the standard temperature of 60 °F (15.56 °C)
5.2 This procedure is most suitable for determining the API gravity of low viscosity transparent liquids This test method can also be used for viscous liquids by allowing sufficient time for the hydrometer to reach temperature equilibrium, and for
3 Available from American Petroleum Institute (API), 200 Massachusetts Ave.
NW, Suite 1100, Washington, DC 20001, http://www.api.org.
4 Available from ASTM International Headquarters Order Adjunct No.
ADJD1250-A1A2-E-PDF Original adjunct produced in 1983.
Trang 3opaque liquids by employing a suitable meniscus correction.
Additionally for both transparent and opaque fluids the
read-ings shall be corrected for the thermal glass expansion effect
before correcting to the reference temperature
5.3 When used in connection with bulk oil measurements,
volume correction errors are minimized by observing the
hydrometer reading at a temperature as close to reference
temperature as feasible
5.4 Gravity is a factor governing the quality of crude oils
However, the gravity of a petroleum product is an uncertain
indication of its quality Correlated with other properties,
gravity can be used to give approximate hydrocarbon
compo-sition and heat of combustion
5.5 Gravity is an important quality indicator for automotive,
aviation and marine fuels, where it affects storage, handling
and combustion
6 Apparatus
6.1 Hydrometers, of glass, graduated in degrees API as
listed inTable 1 and conforming to SpecificationE100or as
listed inTable 2 and conforming to SpecificationE2251
6.1.1 The user should ascertain that the instruments used for
this method conform to the requirements set out above with
respect to materials, dimensions, and scale errors In cases
where the instrument is provided with a calibration certificate
traceable to a NMI (National Metrology Institute), the
instru-ment is classed as certified and the appropriate corrections for
the meniscus effect, the thermal glass expansion effect, and
alternative calibration temperature effects shall be applied to
the observed readings prior to corrections Instruments that
satisfy the requirements of this test method, but are not
provided with a recognized calibration certificate, are classed
as uncertified
6.2 Temperature Determination:
6.2.1 Thermometers (Glass), having a range from −5 °F to
+215 °F and conforming to the requirements for Thermometer
12F as prescribed in SpecificationE1
N OTE 2—The ASTM Gravity Thermometer 12F has 0.5 °F subdivisions
and allowable 60.25 °F scale error and is suitable for use in determining
temperature of bulk crude oil volumes, such as lease production tanks.
Additional thermometers conform to Specification E1 standard or
Speci-fication E2251 standard having a narrower range than the 12F or S12F
Thermometers may also be used, if they have similar performance
characteristics.
6.2.2 Alternate liquid measuring device (Thermometer
S12F) conforming to the requirements prescribed in
Specifi-cationE2251may be used, provided that the total uncertainty
of the calibrated system is no greater than when using
liquid-in-glass thermometers The stated repeatability and re-producibility values are not applicable if alternate fluids are used in the liquid-in-glass thermometers
6.2.3 Digital Temperature Sensors—Digital Contact
Ther-mometers of the PRT style shall meet the requirements of GuideE2877and may be used instead of glass thermometers with the following exceptions:
6.2.3.1 Thermocouples shall not be used
6.2.3.2 Thermistors shall not be used
6.3 Hydrometer Cylinder, transparent material (see 6.3.2) The inside diameter of the cylinder shall be at least 0.25 in (6.35 mm) greater (see A inFig 1) than the outside diameter of the hydrometer body and the temperature measuring device plus the separation interval specified in Fig 1and the height shall be such that the appropriate hydrometer floats in the test portion with at least 1 in (25 mm) clearance between the bottom of the hydrometer and the bottom of the cylinder, under all densities and temperatures Ensure that the hydrometer cylinder is cleaned after each use to ensure that no contami-nants remain
6.3.1 When using separate hydrometer (or thermohydrom-eter) and temperature measuring devices, care shall be exer-cised to ensure that neither interfere with each other and that they are not affected by external temperature effects The minimum separation distances as shown in Fig 1 shall be required These separation distances are a function of the diameters of the measuring devices and the minimum distance from the sides of the measuring chamber to minimize external thermal effects
where:
A = minimum separation interval specified in 6.3 between the inside diameter of hydrometer cylinder and the sum
of the OD of the hydrometer body plus the OD of temperature measuring device plus the minimum sepa-ration interval between devices of 0.25 in (6.35 mm), seeFig 1, and
B = minimum separation interval between measuring de-vices of 0.5 in (12.7 mm)
6.3.2 Hydrometer cylinders constructed of transparent ma-terials shall be resistant to discoloration or attack by the petroleum or petroleum product samples and shall not affect the material being tested They shall not become opaque under prolonged exposure to sunlight If the opacity prevents the
TABLE 1 Available Hydrometers Scaled, Degrees API
Designation Type API Range, deg Scale
Series Total Each Unit Division Error 1H to 10H long plain −1 to 101 12 0.1 0.1
21H to 40H short plain 0 to 101 6 0.1 0.2
41H to 45 H thermo 15 to 51 8 0.1 0.1
51H to 60H thermo −1 to 101 12 0.1 0.1
71H to 74H thermo −1 to 41 12 0.1 0.1
TABLE 2 Available Hydrometers Scaled, Degrees API (Low
Hazardous Liquid Type)
ASTM Hydrometer Designation
Type API Range,
deg Each Unit
Scale Division Error
Trang 4observation of both devices then the hydrometer cylinder shall
be replaced before continuing with the test
6.3.3 The minimum separation intervals specified is to
prevent any capillary action between devices and to ensure that
external temperature effects are minimized on the values
determined for density and temperature
7 Temperature of Test (Limiting Conditions of Test)
7.1 The gravity determined by the hydrometer method is
most accurate at or near the standard temperature of 60 °F
(15.56 °C) Use this or any other temperature between 0 °F and
195 °F (–18 °C and + 90 °C) for the test, so far as it is
consistent with the type of sample and necessary limiting
conditions shown inTable 3
7.2 The purpose of Table 3 is to clarify what actions are
required to ensure that the sample does not change its physical
makeup during the testing period or modifications to the
sample physical properties to ensure that the devices can function properly For example, with highly volatile samples being measured in an open container the light ends may evaporate while the devices are reaching temperature equilib-rium This is a function of ambient temperature as well as fluid temperature and composition Conversely, if the fluid is too viscous (thick) the hydrometer may not float freely, which is a requirement for a buoyance device
8 Procedure
8.1 For referee testing, use the long plain form of hydrom-eter (1H to 10H) For field testing, the thermohydromhydrom-eter method in Test MethodD6822(API MPMS Chapter 9.3) is the
preferred method However, if the user desires to use a liquid-in-glass thermometer with low-hazard glass precision fluid as specified in Specification E2251or a Digital Contact Thermometer as specified in GuideE2877and6.2.3, the user
FIG 1 Separation Intervals (minimum requirements) between Devices and Hydrometer Cylinder
TABLE 3 Limiting Conditions and Testing Temperatures
Sample Type Gravity Limits Initial Boiling Point
Limits Other Limits Test Temperature Highly volatile lighter than 70° API Cool to 35 °F (2 °C) or lower in original closed
container.
Moderately volatile heavier than 70° API below 250 °F (120 °C) Cool to 65 °F (18 °C) or lower in original closed
container.
Moderately volatile and viscous heavier than 70° API below 250 °F (120 °C) Viscosity too high at
65 °F (18 °C)
Heat to minimum temperature for sufficient
fluidity.
Nonvolatile heavier than 70° API above 250 °F (120 °C) Any temperature between 0 °F and 195 °F
(−18 °C and 90 °C) as convenient Mixtures of nonpetroleum products or
essentially pure hydrocarbons
60 °F ± 0.25 °F (15.56 °C ± 0.1 °C)
Trang 5can use Test MethodD6822(API MPMS Chapter 9.3) with a
modified procedure as detailed in 8.11of this test method
8.1.1 As shown inTable 1andTable 2, the user has access
to hydrometers covering various ranges of API gravity The
user should select a hydrometer which results in the liquid
interface cutting the hydrometer stem in the center third of the
range The hydrometer should not be used when the liquid
interface cuts the scale in the bottom two API gravity values or
at the top two API gravity values
8.1.2 When using digital temperature devices the user shall
utilize only intrinsically rated temperature devices that
con-form to the appropriate standard
N OTE 3—In practice whether in a lab or field environment this test
method is used in a potentially hazardous environment (that is, explosive)
and as these devices are electrically powered which could possibly
produce a spark, potentially resulting in an explosion/fire All companies
generally have Engineering and Operating standards that refer to NFPA
(National Fire Protection Association) and NEC (National Electrical
Code) codes that detail the requirements for the use of electrical devices
in classified areas These company requirements should take precedence.
8.2 Prior to lowering the selected hydrometer and/or
tem-perature determination instrument into the sample perform the
Equipment Validation enumerated in Annex A1
8.3 Verify that the selected density and/or temperature
determination instruments conform to the requirements of
Annex A2
8.4 Adjust the temperature of the sample in accordance with
Table 3 For field testing, test temperatures other than those
listed inTable 3may be used The hydrometer cylinder shall be
approximately the same temperature as the sample to be tested
8.5 Transfer the sample into the clean hydrometer cylinder
without splashing, so as to avoid the formation of air bubbles
and to reduce to a minimum the evaporation of the lower
boiling constituents of the more volatile samples (Warning—
Samples may be extremely flammable Vapors may cause flash
fire.) For the more volatile samples, transfer to the hydrometer
cylinder by siphoning (Do not start the siphon by mouth.)
Remove any air bubbles formed, after they have collected on
the surface of the sample, by touching them with a piece of
clean filter paper or other suitable means before inserting the
hydrometer For field testing, the gravity measurement is
directly made in the sampling core thief or hydrometer
cylinder Place the cylinder containing the sample in a vertical
position in a location free from air currents Take precautions
to prevent the temperature of the sample from changing
appreciably during the time necessary to complete the test
During this period, the temperature of the surrounding medium
should not change more than 5 °F (3 °C)
8.6 Lower the hydrometer gently into the sample and, when
it has settled, depress it about two scale divisions into the liquid
and then release it; keep the rest of the stem dry, as unnecessary
liquid on the stem changes the effective weight of the
instrument, and so affects the reading obtained With samples
of low viscosity, a slight spin imparted to the instrument on
releasing assists in bringing it to rest, floating freely away from
the walls of the hydrometer cylinder Allow sufficient time for
the hydrometer to become completely stationary and for all air
bubbles to come to the surface This is particularly necessary in the case of the more viscous samples
8.7 Lower the temperature measuring device slowly into the sample, in close proximity to the hydrometer installed in 8.6
making sure that neither touch, nor come in contact with the side walls of the sample cylinder Refer toFig 2a andFig 2b and Fig 3a and Fig 3b The temperature measuring device may also be used to cautiously and slowly stir the sample instead of the hydrometer, to minimize stratification
8.7.1 The hydrometer is a buoyancy device, its performance
or where it floats is dependent on the temperature of the fluid around the point of buoyancy Therefore, it is essential that the sample temperature be taken as close to this point of buoyancy
In practice with opaque samples this point can be difficult to determine However, it can be estimated from knowing the lengths of the various instruments and then estimating their relative positions
8.8 When the hydrometer has come to rest, floating freely, and the temperature of the sample is constant to within 0.2 °F (0.1 °C), read and record the hydrometer reading to the nearest scale division The correct reading is that point on the hydrometer scale at which the surface of the liquid cuts the scale Determine this point by placing the eye slightly below the level of the liquid and slowly raising it until the surface, first seen as a distorted ellipse, appears to become a straight line cutting the hydrometer scale SeeFig 4
8.9 To make a reading with nontransparent liquids, observe the point on the hydrometer scale to which the sample rises above its main surface, placing the eye slightly above the plane surface of the liquid This reading requires a correction Determine this correction (meniscus correction) for the par-ticular hydrometer in use by observing the height above the
where:
A = separation interval specified in6.3andFig 1,
B = separation interval specified in6.3andFig 1, and
C = relative vertical alignment between the density and temperature measuring devices (the bottom of the sens-ing portion of the temperature measursens-ing device should
be within 60.5 in (612.5 mm) and not more than 61.0 in (625 mm) of the bottom of the density sensor)
FIG 2 a Hydrometer and Liquid-in-Glass Thermometer Placement
(Typical)
Trang 6main surface of the liquid to which the sample rises on the
hydrometer scale when the hydrometer in question is immersed
in a transparent liquid having a surface tension similar to that
of a sample under test SeeFig 5
8.9.1 Record the observed hydrometer scale readings to the
nearest 0.1° API for transparent liquids
8.9.2 When gravity readings have been observed on opaque liquids using the procedure given in8.9, subtract the meniscus correction from the hydrometer reading observed and record the meniscus corrected hydrometer scale reading to the nearest 0.1° API
where:
A = separation interval specified in6.3andFig 1,
B = separation interval specified in6.3andFig 1, and
C = relative vertical alignment between the density and
temperature measuring devices (the bottom of the
sens-ing portion of the temperature measursens-ing device should
be within 60.5 in (612.5 mm) and not more than
61.0 in (625 mm) of the bottom of the density sensor)
FIG 2 b Thermohydrometer and Liquid-in-Glass Thermometer
Placement (Typical) (continued)
where:
A = separation interval specified in6.3andFig 1,
B = separation interval specified in6.3andFig 1, and
C = relative vertical alignment between the density and
temperature measuring devices (the bottom of the
sens-ing portion of the temperature measursens-ing device should
be within 60.5 in (612.5 mm) and not more than
61.0 in (625 mm) of the bottom of the density sensor)
FIG 3 a Hydrometer and Digital Contact Thermometer (Typical)
where:
A = separation interval specified in6.3andFig 1,
B = separation interval specified in6.3andFig 1, and
C = relative vertical alignment between the density and temperature measuring devices (the bottom of the sens-ing portion of the temperature measursens-ing device should
be within 60.5 in (612.5 mm) and not more than 61.0 in (625 mm) of the bottom of the density sensor)
FIG 3 b Thermohydrometer and Digital Contact Thermometer
Placement (Typical) (continued)
FIG 4 Hydrometer Scale Reading for Transparent Liquids
(Typical)
Trang 7N OTE 4—The meniscus correction for a particular hydrometer in use is
determined by observing the maximum height above the principal surface
of the liquid to which liquid rises on the hydrometer scale when the
hydrometer in question is immersed in a transparent liquid having a
surface tension similar to that of the sample under test.
8.10 Observe the temperature of the sample immediately
before and after the observation of the gravity, the liquid in the
cylinder being thoroughly but cautiously stirred with the
thermometer (Note 5), and the whole of the temperature thread
being immersed Should these temperature readings differ by
more than 1 °F (0.5 °C), repeat the temperature and gravity
observations when the temperature of the sample has become
more stable Record the mean of the thermometer reading
before and after the final hydrometer reading, to the nearest
1 °F, as the temperature of the test
8.11 When using a separate temperature measuring device
such as a liquid-in-glass thermometer; a liquid-in-glass
ther-mometer with low-hazard precision liquid or a digital contact
thermometer they shall be placed in the sample measurement
cylinder as shown inFig 2a andFig 2b as well asFig 3a and
Fig 3b
N OTE 5—When only thermohydrometers are used, stir the sample by
carefully moving the thermohydrometer in a side to side motion, without
immersing the thermohydrometer any lower into sample It is satisfactory
in this case to read the thermometer scale of the thermohydrometer after
the hydrometer reading has been observed Read the thermometer to the
nearest 1 °F (0.5 °C) If using a separate device to measure temperature,
use the temperature measuring device to stir the sample and read
according to 8.10
9 Calculation
9.1 Apply any relevant thermometer corrections to the temperature reading observed in8.10and record the average of those two temperatures to the nearest 1 °F
9.2 Application of the glass thermal expansion correction depends upon what edition of Adjunct to D1250 Standard Guide for the Use of the Joint API and ASTM Adjunct for Temperature and Pressure Volume Correction Factors for Generalized Crude Oils, Refined Products, and Lubricating
Oils (API MPMS Chapter 11.1) will be used to calculate the
base density
9.2.1 The 1980 version of the Adjunct toD1250Guide for
Petroleum Measurement Tables (API MPMS Chapter 11.1) has
the hydrometer glass thermal expansion correction included Input into the Adjunct toD1250Standard Guide for the Use of the Joint API and ASTM Adjunct for Temperature and Pressure Volume Correction Factors for Generalized Crude Oils,
Re-fined Products, and Lubricating Oils (API MPMS Chapter 11.1) software (see API MPMS Ch 11.1.1.3 paragraphs 5 and
6) requires the Hydrometer Reading – Observed or Hydrometer Reading – Observed, Meniscus Corrected in API units from
8.9.1 or 8.9.2, observed temperature of the sample, and the built-in hydrometer glass thermal correction switch set to “on” (0) or “off” (1) It will return API @ 60 °F
9.2.2 The 2004 version of the Adjunct toD1250Guide for
Petroleum Measurement Tables (API MPMS Chapter 11.1)
does not include the hydrometer glass thermal expansion correction, so that correction must be made before entering the Adjunct toD1250Standard Guide for the Use of the Joint API and ASTM Adjunct for Temperature and Pressure Volume Correction Factors for Generalized Crude Oils, Refined
Products, and Lubricating Oils (API MPMS Chapter 11.1)
software Depending on the specific end use of the calculation results, the final value may be left rounded or unrounded See
9.3 9.3 The following steps are required to implement9.2.2:
Step 1. Convert the meniscus corrected hydrometer scale reading to density in kg/m3using Eq 2
Hydrometer Scale Reading Units
Conversion to Density For API gravity:
density~kg / m3!5~141.5*999.016!/~131.51API! (2)
Leave the result unrounded
Step 2 Calculate the hydrometer thermal glass expansion
correction factor (HYC) using the appropriate equation below
(t is observed temperature).
Correction for a Base Temperature (T b) of 60 °F:
HYC 5 1.0 2@0.00001278~τ 2 60!#2@0.0000000062~τ 2 60!2#
(3)
Leave the result unrounded
Step 3 Multiply the hydrometer reading in kg/m3from Step
1 by HYC from Step 2 to obtain the glass expansion corrected hydrometer reading
kg/m3
FIG 5 Hydrometer Scale Reading for Opaque Fluids (Typical)
Trang 8Step 4a Convert the hydrometer reading in density (kg/
m3HYC) from Step 3 to a R.D (relative density) hydrometer
reading
N OTE 6—The current C source code, compiled dll and Excel Add-in has
an omission and cannot use a kg/m 3 call with degree F.
R.D 5 kg/m3
Step 4b Input R.D and degree F into section 11.1.6.2 of the
Adjunct to D1250-04 Guide for Petroleum Measurement
Tables (API MPMS Chapter 11.1-2004) which returns R.D @
60 °F
N OTE 7—Pressure will have to be atmospheric gauge, or 0 psig as the
Adjunct to D1250 Standard Guide for the Use of the Joint API and ASTM
Adjunct for Temperature and Pressure Volume Correction Factors for
Generalized Crude Oils, Refined Products, and Lubricating Oils (API
MPMS Chapter 11.1) values are only valid at atmospheric pressure.
Step 4c Convert the calculated R.D value @ 60 °F to a
calculated API Gravity @ 60 °F usingEq 6
9.4 Future versions of the Adjunct toD1250Standard Guide
for the Use of the Joint API and ASTM Adjunct for
Tempera-ture and Pressure Volume Correction Factors for Generalized
Crude Oils, Refined Products, and Lubricating Oils (API
MPMS Chapter 11.1) code will be corrected so that it can
accept any combination of input units and return any
combi-nation of output units When available, the Adjunct to D1250
Standard Guide for the Use of the Joint API and ASTM
Adjunct for Temperature and Pressure Volume Correction
Factors for Generalized Crude Oils, Refined Products, and
Lubricating Oils (API MPMS Chapter 11.1) code can be
accessed directly from Step 3 and return API Gravity @ 60 °F,
R.D @ 60 °F, and kg/m3at any selected base temperature
Example 1
Observed Temperature: 77 °F
Observed Hydrometer Reading: 33.2 API Gravity
Observed Pressure: 0 psig
Base Temperature: 60 °F
Step 4c.1: 31.955643312 unrounded ( Eq 6 )
Step 4c.2: 32.0 °API rounded ( Eq 6 )
10 Report
10.1 Report the corrected hydrometer reading as degrees
API (°API) or as API Gravity
10.2 Report the final value as API gravity, at the reference temperature (nearest whole degree °F), to the nearest 0.1° API 10.3 The reporting values have no precision or bias deter-mination It is up to the user to determine whether this test method provides results of sufficient accuracy for the intended purpose
10.4 If the hydrometer readings are being used as an input to
a calculation process intended to return a volume correction factor for use in ticket or meter proving calculations, stop the calculation process identified above at Step 3 (if the density value is desired at flowing conditions) or Step 4 (if the density value is desired at base density conditions) and input the results into the calculation process
10.5 Certified hydrometers traceable to a NMI (National Metrology Institute), report the output density as ‘Density in Vacuo’
11 Precision and Bias
11.1 The precision of this test method as obtained by statistical examination of interlaboratory test results is as follows:
11.1.1 Repeatability—The difference between successive
test results obtained by the same operator with the same apparatus under constant operating conditions on identical test material, would in the long run, in the normal and correct operation of the test method, exceed 0.2° API only in one case
in twenty
11.1.2 Reproducibility—The difference between two single
and independent results, obtained by different operators, work-ing in different laboratories on identical test material, would in the long run, in the normal and correct operation of the test method, exceed 0.5° API only in one case in twenty
N OTE 8—The precision for this test method was not obtained in accordance with RR:D02-1007.
N OTE 9—The precision for this test method only applies to measure-ments made with a liquid-in-glass thermometer.
N OTE 10—This precision statement applies only to measurements made
at temperatures differing from 60 °F (15.56 °C) by less than 18 °F (10 °C).
11.2 Bias—Bias for this test method has not been
deter-mined
12 Keywords
12.1 API gravity; crude petroleum; digital contact thermom-eter; hydromthermom-eter; thermohydromthermom-eter; thermometer
Trang 9ANNEXES (Mandatory Information) A1 PERFORMANCE CHECK EQUIPMENT VALIDATION AS TO FIT FOR USE AT TIME OF TEST
A1.1 Immediately prior to determining test values the
den-sity and temperature measuring instruments being used for the
test shall be verified that they are “Fit for Use at Time of Test”
as delineated by the following procedures The specific
proce-dures the user shall use will be dependent on the specific test
instruments being used for that test
A1.2 These verifications shall be performed on each test
instrument immediately prior to each test being performed
A1.3 Each test instrument shall be in a clean condition prior
to the equipment validation so that each component of the Test
Instrument can be validated
A1.4 Equipment Validation Procedures
A1.4.1 Density Instruments:
A1.4.1.1 Hydrometers:
(1) Ensure that the hydrometer scale range selected for the
analytical test covers the expected density reading
(2) Inspect the hydrometer and ensure that it meets the
following requirements:
(a) The hydrometer glass shall contain no cracks, fissures,
deep scratches, rough areas, or other obvious damage
(b) The ballast at the bottom of the hydrometer shall
contain no loose components
(c) The ballast at the bottom of the hydrometer shall be
firmly affixed to glass envelope of the test instrument
(d) The paper scale within the hydrometer stem shall be
straight and without twist
(e) The paper scale within the hydrometer stem shall
contain an information label that documents the ASTM
hy-drometer designation, units of density, and a serial number
(f) The hydrometer shall contain a scale slippage
indica-tor and shall be positioned at the proper position relative to the
scale This is typically the whole unit value graduation mark
for the upper range the instrument is designed for
(3) If any of the items in A1.4.1.1(2) do not meet the
defined criteria, the instrument shall no longer be used and
shall be replaced
A1.4.1.2 Thermohydrometers:
(1) Ensure that the thermohydrometer scale range selected
for the analytical test covers the expected density reading
(2) Inspect the thermohydrometer and ensure that it meets
the following requirements:
(a) The thermohydrometer glass shall contain no cracks,
fissures, deep scratches, rough areas, or other obvious damage
(b) The ballast at the bottom of the thermohydrometer
shall contain no loose components
(c) The ballast at the bottom of the thermohydrometer
shall be firmly affixed to glass envelope of the test instrument
(d) The paper scale within the thermohydrometer stem
shall be straight and without twist
(e) The paper scale within the thermohydrometer stem
shall contain an information label that documents the ASTM hydrometer designation, units of density, and a serial number
(f) The thermohydrometer shall contain a scale slippage
indicator and shall be positioned at the proper position relative
to the scale This is typically the whole unit value graduation mark for the upper range the instrument is designed for
(3) If any of the items in A1.4.1.2(2) do not meet the
defined criteria, the instrument shall no longer be used and shall be replaced
A1.4.2 Temperature Instruments:
A1.4.2.1 Liquid-in-Glass Type Instruments:
(1) Ensure that the liquid-in-glass temperature instrument
scale range selected for the analytical test covers the expected temperature reading
(2) Inspect the liquid-in-glass temperature instrument to
ensure it meets the following requirements:
(a) The temperature instrument glass shall contain no
cracks, fissures, deep scratches, rough areas, or other obvious damage
(b) The temperature glass instrument shall contain
per-manent markings that document the ASTM temperature instru-ment designation and the instruinstru-ment’s serial number
(c) The liquid column is continuous from the fluid
reser-voir to the indicated temperature
(d) Ensure that the temperature scale and its graduation
markings are intact on the instrument
(3) If any of the items in A1.4.2.1(2) do not meet the
defined criteria, the instrument shall no longer be used and shall be replaced, with the exception of itemA1.4.2.1(2)(c) (4) If the liquid column is non-continuous, the temperature
instrument shall be placed in a liquid bath and the temperature
of the bath increased until it causes the liquid column to expand into the upper expansion area After this occurs, remove the temperature instrument from the liquid bath and cool the instrument down to ambient temperature Reverify that the liquid column is now continuous
(5) If the liquid column is now continuous verify that the
instrument still indicates the correct temperature in accordance
wiAnnex A2th If the column is non-continuous the instrument shall not be used, perform the procedure in A1.4.2.1(4) again
to try to re-unite the column and then re-verify perA1.4.2.1(5) A1.4.2.2 Digital Contact Thermometer:
(1) Inspect the Digital Contact Thermometer to ensure it
meets the following requirements:
(a) The temperature instrument sensor shall be a PRT
style meeting the requirements of GuideE2877
(b) The temperature instrument shall meet the
require-ments of 6.2.3of this test method
(c) Inspect the temperature instrument and ensure that all
components meet all mechanical and electrical requirements
Trang 10(d) Perform an operational performance check in
accor-dance with the manufacturer’s requirements:
(1) Note—For digital contact thermometers being used
multiple times in a short period of time, such as determining
the temperature reading of multiple samples in a laboratory
over a very short interval (an hour or less) this step may be
skipped once it is performed once
(2) However, if an interval of 60 min or more is
experi-enced between analytical tests then this performance check is
required prior to any further testing
(3) Operational Performance Check:
(a) Verify that the device is in good working order.
(b) Verify that there is no mechanical damage to the
instrument case body, temperature sensor, or the connecting
wiring
(c) Verify that the battery voltage is in the manufacturer’s
recommended operating range
(d) Verify that the display(s) are legible.
(e) Verify that when turned on the temperature reading is
indicating a change and the change is indicating in the correct direction:
(1) The comparison device for the operational
perfor-mance check could be a liquid-in-glass thermometer; a pocket dial thermometer; an ambient air reading from a local source or other suitable source
(2) This check is intended to verify that the instrument is
working properly and not that it is measuring temperature accurately
(e) Verify that the temperature instrument shall have been
calibrated in accordance withAnnex A3
(2) The digital contact thermometer shall be calibrated in
accordance with Annex A3 if any of the requirements in
A1.4.2.2are not met prior to any use
A2 PERFORMANCE VERIFICATION OF DENSITY AND TEMPERATURE MEASURING INSTRUMENTS
A2.1 All density and temperature measuring instruments
shall be verified on a periodic schedule, which depends on the
frequency of use
A2.2 Verification Process—Verification is a process
whereby the working instrument is compared to another similar
device with a higher accuracy (lower uncertainty) This device
is typically referred to as a “Primary Verification Device or
Master Device”
A2.2.1 The Primary Verification Device shall have been
calibrated using standards traceable to a NMI (National
Me-trology Institute) and calibrated in accordance with the
appro-priate sections of Test MethodE126
N OTE A2.1—In the United States, the NMI is the National Institute of
Standards and Technology (NIST) and in Canada it is Measurement
Canada Many companies which supply hydrometers, thermometers to the
industry have the capability to provide equipment with NMI traceability.
A2.2.2 The calibration of the Primary Verification Device
shall be performed on an annual basis
A2.2.3 Verifications shall be performed using a liquid bath
both for density and temperature instruments
A2.2.3.1 A Primary Verification Device used with a test/dry
block may be substituted for a liquid bath for temperature
instruments, provided that it facilitates the ability to determine
the temperature readings of both the Primary Verification
Device and the working instrument without moving the
instru-ments within the test block
A2.3 Verification Test Points:
A2.3.1 Density Instruments:
A2.3.1.1 Test Method E126shall be used as the reference standard (for the calibration of the Primary Verification De-vice) for the verification of density instruments
A2.3.1.2 Density instruments shall be verified at two test points (approximately lower half and upper half on the hydrometer scale) on the density scale Test points shall be chosen to bracket the range of use
A2.3.1.3 For routine verification work, one may use a standard hydrometer as specified in Section 6 of Test Method
E126 and a reference fluid with similar surface tension and density of the liquids the hydrometer is intended to be used, see Table 2 in Test Method E126for guidance
A2.3.2 Temperature Instruments:
A2.3.2.1 Test Method E77 shall be used as the reference standard for the verification of temperature instruments
(1) Secondary standard thermometers as specified in 5.5 of
Test Method E77are more suitable for routine work
(2) Alternatively the user may also use a primary standard
thermometer as specified in 5.4 of Test MethodE77 A2.3.2.2 Temperature instruments shall be verified at two or more test points on the temperature scale Test points shall be chosen to bracket the range of use
A2.4 Frequency of Verification:
A2.4.1 Density Instruments:
A2.4.1.1 Hydrometers and thermohydrometers should be verified monthly (at a minimum they shall be verified quar-terly) for frequently used instruments (used four or more times
in a seven day period)
A2.4.1.2 Hydrometers and thermohydrometers should be verified bi-monthly (at a minimum they shall be verified every